Radiation Dose Can Be Reduced For “Triple Rule-Out” Coronary CT Angiography

Physicians can dramatically reduce the radiation dose delivered to patients undergoing coronary CT angiography in a “triple rule-out” protocol by simply using tube current modulation, according to a study performed at Thomas Jefferson University Hospital in Philadelphia, PA.

The study included 172 patients who were evaluated using coronary CT angiography without tube current modulation and 95 patients who were evaluated with tube current modulation. The effective radiation dose ranged between 9.9 and 31.3 mSv in patients without modulation; the dose ranged between 5.4 and 16.6 mSv in patients with modulation. “Image quality was comparable to when we didn’t use tube current modulation,” said Kevin M. Takakuwa, MD, lead author of the study.

“One of the major criticisms of the triple rule out coronary CT angiography study is the concern about the high amount of radiation given, which has been estimated by some to be as high as 30-40 mSv. Our study demonstrates that the radiation is a lot less, averaging less than 9 mSv when using tube current modulation. In addition, it uses less radiation than a nuclear stress test, a common alternative study to the triple rule-out cardiac CT,” he said.

source: ARRS